Known Limitations

This document lists the known limitations and edge cases of the current OpenARaw reader that result in unparseable data or conformance invariant violations across the PRIDE corpus.

1. Malformed or Corrupted Folders​

Some files uploaded to PRIDE are structurally malformed, missing the mandatory AcqData directory entirely.

Affected Files (6 files):

  • PXD041903: 20190423_Alex11.d, 20190423_Alex3.d, 20190423_Alex7.d (Double-nested folders missing AcqData in the top level).
  • PXD049393: __MACOSX/2022-07-07_sequence1.d, __MACOSX/2023-03-01_sequence1.d, __MACOSX/2023-07-12_sequence1.d (macOS filesystem artifacts rather than real datasets).

2. selected_mz Not Populated From MSScan.bin​

PrecursorInfo::selected_mz is always None in the reader's output; no source for it is known:

  • selected_mz: distinct from target_mz (the isolation window center, which the reader already exposes via PrecursorInfo::target_mz). No f32/f64 offset in the record holds a second m/z-like value near target_mz across hundreds of MS2 scans spanning strides 216, 220, and 284. Either Agilent does not store a distinct selected/monoisotopic precursor m/z in this record (recovering one would require re-picking against the isolation window in MSPeak.bin, a processing step this reader does not perform), or it is identical to target_mz and therefore not a distinct field.

A source for selected_mz outside MSScan.bin (e.g. a fuller decode of MSPeriodicActuals.bin's per-channel Value semantics) is unchecked. Any change here needs corpus evidence, not just a plausible byte offset.

3. No Ion Mobility / CCS Support (Agilent 6560 IM-QTOF)​

SpectrumRecord::inv_mobility and SpectrumRecord::inv_mobility_per_peak are always None; RunMetadata::mobility_array_kind is always None as well. Unlike section 2, there is no real record to check against: no ion-mobility acquisition exists anywhere in the validation corpus.

All 338 cataloged .d directories come from the PRIDE Archive (see CORPUS.md), a proteomics-focused repository. Every Devices.xml mass spectrometer <ModelNumber> seen across the corpus is one of G6410A (QQQ), or G6530A/G6530B/G6540A/G6540B/G6550A/G6550B (Q-TOF) - see docs/format/01-msscan.md and docs/format/07-run-metadata.md. No G6560 (6560 IM-QTOF) unit, or any other drift-tube-equipped instrument, appears in any cataloged bundle.

Without a real 6560 AcqData bundle to read, there is nothing to reverse-engineer against - and per this project's clean-room policy, guessing at a plausible-looking field or file name from general IM-MS knowledge and shipping it as a "decode" is exactly the kind of unverified claim the policy prohibits. This is therefore left undocumented-as-a-format-question rather than implemented:

  • It is not confirmed whether a 6560 AcqData directory even reuses MSScan.bin's existing record layout (e.g. an extra drift-time field appended to the strides already catalogued in docs/format/01-msscan.md), or whether IM-QTOF acquisitions use additional binary stream(s) alongside it (frame-indexed drift spectra, the way OpenWRaw and OpenTimsTDF each expose an instrument-specific mobility file distinct from their base spectrum index) - this crate has never seen one to check.
  • It is not confirmed what such a file would be named, or what its header/record layout would be, without a sample to inspect.
  • CCS (collision cross section) is typically a derived value (from drift time, charge state, and instrument calibration constants) rather than a directly stored field in most vendor IM formats; whether Agilent stores a precomputed CCS anywhere in the bundle, or only raw drift time requiring the reader to apply a calibration, is unknown without a sample.

Prerequisite for support: obtain a real 6560 IM-QTOF .d bundle from a public, non-vendor-tool-verified source (a PRIDE, MetaboLights, or similar open-data submission using this instrument) and add it to the corpus before writing any parsing code. Without one, inv_mobility / inv_mobility_per_peak / mobility_array_kind should stay None rather than be backed by an unverified field guess.

4. Q-TOF Calibration Files and Missing Profile Data​

  • Q-TOF bundles with neither MSMassCal.bin nor DefaultMassCal.xml cannot be decoded. Without either file there is no calibration to convert flight times to m/z. iter_spectra yields an error for each scan (counted in Reader::profile_decode_failures() / Reader::centroid_decode_failures()); flight times are never reported as m/z. One corpus bundle is affected: PXD031771/655_2.d, which also lacks Devices.xml and the other method files.
  • PXD022473/M.d has no MSProfile.bin although its scan records point into one (profile data only). Every scan is yielded as an error and counted as a profile decode failure. This is a defect of the upload, not a format variant.
  • QQQ (MRM) centroid data stores m/z directly and is unaffected.