Q-TOF mass calibration. Q-TOF flight times (MSPeak.bin FormatID=2 x-values, the MSProfile.bin axis) are converted to m/z with a two-step calibration. MSMassCal.bin holds one calibration per scan; DefaultMassCal.xml holds the defaults and the Polynomial step's ValueUseFlags, which the binary records do not store. QQQ bundles have neither file.
Status: Confirmed
The formula below reproduces the BasePeakMZ stored in every Q-TOF MSScan.bin record to well under 0.001 ppm across the corpus (see Corpus Evidence).
Calibration Formula
m/z = (A * (t - T0))^2 - sum_k c_k * clamp(t, t_min, t_max)^k
tis the flight time as stored (centroid x-value, orx0 + i * dxon the profile axis).Traditionalstep:AandT0.Polynomialstep: eight values[t_min, t_max, c_0, ..., c_5]. The polynomial is evaluated attclamped to[t_min, t_max], and its result is subtracted from theTraditionalm/z.- Powers: the set bits of the
Polynomialstep'sValueUseFlags, lowest bit first. Coefficient slotjbelongs to thej-th set bit; slots beyond the number of set bits are zero. Example:ValueUseFlags=232(bits 3, 5, 6, 7) meansc_0*t^3 + c_1*t^5 + c_2*t^6 + c_3*t^7. The 37 distinct flag values in the corpus select between 0 and 6 powers in the range 0-11.
MSMassCal.bin Layout
- Global header: 72 bytes. In all 188 corpus files: int32
260(0x04 0x01 0x00 0x00) at offset 0, int321at offset 68, zeros elsewhere. Not otherwise interpreted. - Record stride: 84 bytes.
(FileSize - 72)is an exact multiple of 84. - Record count: equals the number of scans in
MSScan.binin all 188 files; recordibelongs to scan recordi.
Record Layout (84 bytes, little-endian)
| Bytes | Field | Type |
|---|---|---|
| 0-3 | NumCoefficients | int32 (10 in every corpus record) |
| 4-83 | Coefficients | double[10]: A, T0, then the 8 Polynomial values |
DefaultMassCal.xml Layout
UTF-8 with a byte-order mark. All 222 corpus files have the same structure: two DefaultCalibration entries (DefaultCalibrationID 1 and 2), each with a Traditional step (NumberOfCoefficients=2, ValueUseFlags=0) and a Polynomial step (NumberOfCoefficients=8). CalibrationTechnique is ExternalReference throughout. Values are decimal text with 15 significant digits, so they equal the binary doubles to about 1e-15 relative.
MSScan.bin CalibrationID (record offset 52) names the DefaultCalibrationID a scan uses.
Per-Scan Records and Reference-Mass Correction
In every corpus record, the 8 Polynomial values equal those of the DefaultMassCal.xml entry named by the scan's CalibrationID. A, and in some bundles T0, differ from that entry: the instrument adjusts them per scan from the infused reference-mass ions. Scans the instrument did not adjust keep the default A and T0.
The stored BasePeakMZ is computed with the DefaultMassCal.xml calibration, before this adjustment. Decoding with the per-scan record therefore differs from BasePeakMZ by the adjustment, and puts the reference ions at their theoretical m/z where the instrument applied it (see Corpus Evidence).
Reader Behaviour
crates/openaraw/src/raw/calibration.rs, defaultmasscal.rs and Reader::open:
- Each scan uses its own
MSMassCal.binrecord. TheValueUseFlagscome from theDefaultMassCal.xmlentry whosePolynomialvalues equal the record's to within 1e-12 relative, preferring the entry named by the scan'sCalibrationID. A record whosePolynomialcoefficient slots are all zero needs no flags. - When
MSMassCal.binis absent, each scan uses theDefaultMassCal.xmlentry for itsCalibrationID(logged at info level). This is normal: about a sixth of the corpus Q-TOF bundles have noMSMassCal.bin. - When
MSMassCal.binis unreadable, its record count differs from the scan count, or a record is unusable, the affected scans useDefaultMassCal.xmlinstead, with a warning. - A scan with no usable calibration from either source gets none; its payloads are logged and counted as decode failures, never reported as uncalibrated flight times.
Reader::calibration(i)returns the calibration and itsCalibrationSource.
Corpus Evidence
Measured with cargo run --release --example qtof_calibration_check -- <bundle.d> on all 224 Q-TOF bundles of the research corpus (24 PRIDE accessions; models G6530A/B, G6540A/B, G6550A/B; acquisition software B.05 to 10.1):
- Formula: in the 209 bundles whose spectra decode (5.4 million scans), decoding the base peak with the
DefaultMassCal.xmlcalibration reproduces the storedBasePeakMZwith a worst-bundle median of 0.00012 ppm and a worst-bundle p95 of 0.00049 ppm. 174 of these bundles useMSMassCal.binrecords, 35 onlyDefaultMassCal.xml. In the 13PXD044639bundles (run-length profile blocks, 192,503 scans; see04-msprofile.md) the profile base peak matchesBasePeakMZwithin 0.0002 ppm on every scan. - What
BasePeakMZdescribes: in 185 bundles the storedTICequals the summed centroid intensity andBasePeakMZis the m/z of the most intense centroid; in 24 profile bundles (acquisition software B.06 to B.09) and the 13PXD044639bundles (10.1)TICequals the summed profile intensity,BasePeakValuethe most intense profile point, andBasePeakMZthat point's m/z. PXD001310 (B.05.01) has profile data but centroid-based summary fields. - Without the
Polynomialstep (Traditionalonly) the per-bundle median error is 0.56 to 33 ppm. Evaluating the polynomial without clamping to[t_min, t_max]leaves about 1 ppm on scans outside the range (PXD004426). - Reference-mass correction: with the per-scan records, the lock mass sits within 0.1 ppm (median) of its theoretical m/z in 58 of the 60 bundles where HP-0921 (
922.009798) is found and 53 of the 68 where HP-1221 (1221.990637) is found; PXD001310 HP-1221: 0.004 ppm, against about -10 ppm with the default calibration. In the remaining bundles most records keep the defaultA, i.e. the instrument did not adjust those scans. - Not decodable:
PXD031771/655_2.dhas neither calibration file;PXD022473/M.dhas noMSProfile.binfor its profile-only scans (see06-known-limitations.md). - QQQ bundles (PXD004747, PXD028295) have neither file.
Unresolved
- The meaning of the non-zero header words (offsets 0 and 68).