MSPeriodicActuals.bin

Binary telemetry log. Always present in .d/AcqData/. Records instrument diagnostic values (pressures, temperatures, voltages, charge states, etc.) at irregular intervals throughout the run. Field definitions are in the companion MSActualDefs.xml.

Status: Confirmed

File Layout

  • Global header: 68 bytes.
  • Record stride: 20 bytes (confirmed via file size arithmetic and sequential parsing on PXD004747 and PXD001310).

Record Layout (20 bytes, little-endian)

The layout has two variants depending on whether the instrument is a QQQ or Q-TOF. The ActualID and Time fields are always at the same offsets; the Value field position shifts.

QQQ variant (PXD004747, G6410A)

BytesFieldType
0-3OccurrenceFlagint32 (1 on first event for this ActualID in a block, else 0)
4-7ActualIDint32
8-15Timedouble (minutes)
16-19Valuefloat32 or int32 (see DataType)

Q-TOF variant (PXD001310, G6550A)

BytesFieldType
0-3Valuefloat32 or int32 (see DataType)
4-7ActualIDint32
8-15Timedouble (minutes)
16-19Paddingint32 (always 0, or int32 for DataType=3)

DataType Mapping (from MSActualDefs.xml)

All DataType codes observed across QQQ (PXD004747) and Q-TOF (PXD001310) files:

DataTypeValue TypeValue Offset (QQQ)Value Offset (Q-TOF)Examples
3int32offset 16offset 16Charge State (ActualID=57), Min Range (ActualID=361)
5float32offset 16offset 0Capillary Current (QQQ, ID=10), Funnel DC (Q-TOF, ID=85)
6float32N/A (not in QQQ)offset 0Rough Vac (ID=29, ~0.52 Torr), Quad Vac (ID=30), Gas Temp (ID=27), Drying Gas (ID=34)
  • DataType=5 QQQ example: ActualID=10 (Capillary Current), value ~39 uA at offset 16.
  • DataType=5 Q-TOF example: ActualID=85 (Funnel DC), value 4.76 V at offset 0.
  • DataType=6 example: ActualID=29 (Rough Vac), value 0.52 Torr at offset 0. Physically plausible for a rough vacuum pump.
  • DataType=3 example: ActualID=57 (Charge State), int values 0, 2, 3, 5, 7 at offset 16 (Q-TOF).

Unresolved

  • Whether the QQQ/Q-TOF variant distinction is instrument-model-specific or encoded somewhere in the file header.
  • DataType values beyond 3, 5, 6 are not observed in the two files checked. The full set of DataType codes in the Agilent schema is not known from the XML alone.